Biosafety Cabinet Manufacturers in Chennai

We are a leading  Biosafety Cabinet Manufacturers in Chennai, All model Biosafety Cabinets such as Class II Type A1, Class II Type A2, Class II Type B1, Class II Type B2, Class III are available.
1, Biosafety Cabinet – Class II Type A1
“Pureair” Biosafety Cabinet Class II Type A1 is designed for protection of operator, product and environment from pathogenic agents and microorganisms transmitted in respiratory way and also for creation of a bacterial dust free air environment. It has a common plenum from which 30% of air is exhausted, and 70% re-circulated to the work area as the downflow. Type A cabinets exhaust air directly back to the laboratory, and they may contain positive pressure contaminated plenums. When toxic chemicals must be employed as an adjunct to microbiological processes, these cabinets should not be used. Exhaust HEPA filtration only removes airborne aerosols including biohazards, and not chemical fumes.

The Class II Type A1 has the positively-pressurized contaminated plenum bordering the ambient environment, and therefore is less safe than the Class II Type A2 that has a negative pressure surrounding the positively pressurized contaminated plenum. In case there is a leakage on the positive plenum, the leaking aerosol will be pulled by the negative pressure back to the positive plenum, and it will not leak out. Because of the safety issue, the Type A1 design is now considered obsolete. In the A1 cabinet, about 70% of air from the positive plenum is recirculated as downflow, and the remaining All Class II biological safety cabinets type A1 is suitable for work with microbiological agents assigned to biosafety levels 1, 2 and 3.

Working Area – Stainless Steel IS 304 Grade Satin Finish

Product Biosafety Cabinet – Class II- Type A1
Air balancing 70% recirculation, 30% exhaust
Protection Product, Personnel & Environmental protection
Cleanliness Level ISO Class 5 as per ISO 14644-1
Average Airflow Inflow Velocity: Minimum 0.45 m/s (90 FPM) at 8” front opening
Downflow Velocity: 0.30 m/s (65 FPM)
Noise Level 60 dB on A scale
Light Intensity Minimum 900 lux inside the working area at ambient zero
MOC Main Body – Galvanized Iron with PU coated / Stainless Steel – Satin Finish
Working Area – Stainless Steel IS 304 Grade Satin Finish
Supply Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency: 99.99% down to 0.3 Microns.
Exhaust Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency: 99.99% down to 0.3 Microns.
Pre Filter “Pureair” Make – EU 4 Rating with suitable size.
Efficiency: 90% down to 10 Microns.
Noise Level Less than 70 dB ‘A’
Vibration Level Minimum
Magnehelic Differential Pressure Gauge 01 No. (Make: Dwyer, USA) – To measure differential pressure across the supply HEPA filter.
Motor Blower Assembly Statically & Dynamically balanced with suitable rating and size.
Induction motor with centrifugal blower.
Mounting: Unique spring suspension system to reduce vibration & noise.
3 Pin Switch & Socket 15 AMPS – 1 set in each unit
Other Switches On/Off switches for motor blower & CFL lights

2, Biosafety Cabinet – Class II Type A2

“Pureair” Biosafety Cabinet Class II Type A2 is designed for protection of operator, product and environment from pathogenic agents and microorganisms transmitted in respiratory way and also for creation of a bacterial dust free air environment. It has a common plenum from which 30% of air is exhausted, and 70% re-circulated to the work area as the down flow. Type A cabinets exhaust air directly back to the laboratory, and they may contain positive pressure contaminated plenums. When toxic chemicals must be employed as an adjunct to microbiological processes, these cabinets should not be used. Exhaust HEPA filtration only removes airborne aerosols including biohazards, and not chemical fumes.

The Class II Type A1 has the positively-pressurized contaminated plenum bordering the ambient environment, and therefore is less safe than the Class II Type A2 that has a negative pressure surrounding the positively pressurized contaminated plenum. In case there is a leakage on the positive plenum, the leaking aerosol will be pulled by the negative pressure back to the positive plenum, and it will not leak out. Because of the safety issue, the Type A1 design is now considered obsolete. In the A2 cabinet, about 70% of air from the positive plenum is recirculated as down flow, and the remaining All Class II biological safety cabinets A2 is suitable for work with microbiological agents assigned to biosafety levels 1, 2 and 3.

No. Property Details
1ProductClass II Biosafety Cabinet – Type A2
2Air balancing70% recirculation, 30% exhaust
3ProtectionProduct, Personnel & Environmental Protection
4Bio safety LevelBio Safety Level 1, 2 & 3
5Exhaust RequirementReturn to Laboratory or Thimble Ducting Only
6Cleanliness LevelISO Class 5 as per ISO 14644-1
7Average AirflowInflow Velocity Minimum 0.45 m/s (90 FPM) at 8” front opening
8VelocityDown Velocity 0.30 m/s (65 FPM)
9Noise Level60 dB on A scale
10Light IntensityMinimum 900 lux inside the working area at ambient zero
11 MOC Main Body – Galvanized Iron with PU coated / Stainless Steel – Satin Finish
Working Area – Stainless Steel IS 304 Grade Satin Finish
12 Supply Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency 99.99% down to 0.3 Microns.
13 Exhaust Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency 99.99% down to 0.3 Microns.
14 Pre Filter “Pureair” make, EU 4 rating with suitable size.
Efficiency 90% down to 10 Microns.
15Noise LevelLess than 70 dB ‘A’
16Vibration Level / InstrumentationMinimum
17 Magnehelic Differential Pressure Gauge 01 No. (Make: Dwyer, USA) with each unit
(To measure differential pressure across the supply HEPA filter).
18 Motor Blower Assembly Statically & Dynamically balanced with suitable rating and size.
(Induction motor with centrifugal blower)
Mounted on unique spring suspension system to reduce vibration and noise.
193 Pins Switch & Socket15 AMPS – 1 set in each unit
20Other SwitchesOn/Off switches for motor blower & CFL lights
PAH 600 PAH 900 PAH 1200 PAH 1800
Horizontal PAH 600 PAH 900 PAH 1200 PAH 1800
Vertical PAV 600 PAV 900 PAV 1200 PAV 1800
Working Size 2 x 2 x 2 3 x 2 x 2 4 x 2 x 2 6 x 2 x 2

3, Biosafety Cabinet – Class II Type B1

“Pureair” Biosafety Cabinet Class II Type B1 has a common plenum from which 70% of air is exhausted, and 30% re-circulated to the work area as the down flow.

Type B1 cabinets also have a dedicated exhaust feature that eliminates re-circulation when work is performed towards the back within the interior of the cabinet. Toxic chemicals employed as an adjunct to microbiological processes should only be employed if they do not interfere with work when re-circulated in the down flow.

No. Property Details
1ProductBio Safety Cabinet – Class II Type B1
2Air Balancing30% recirculation, 70% exhaust
3ProtectionProduct, Personnel & Environmental protection
4Bio Safety LevelBio Safety Level 1, 2 & 3
5Exhaust Duct RequirementHard Duct Only
6Control Plenum Surrounded ByNegative Plenum
7Cleanliness LevelISO Class 5 as per ISO 14644-1
8 Average Airflow / Velocity Inflow Velocity – Minimum 0.45 m/s (90 FPM) at 8” front opening
Downflow Velocity – 0.30 m/s (65 FPM)
9Noise Level60 dB on A scale
10Light IntensityMinimum 900 lux inside the working area at ambient zero
11 MOC Main Body – Galvanized Iron with PU coated / Stainless Steel – Satin Finish
Working Area – Stainless Steel IS 304 Grade Satin Finish
12 Supply Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency: 99.99% down to 0.3 Microns.
13 Exhaust Filter HEPA Filter having EU 13 rating with suitable size.
Efficiency: 99.99% down to 0.3 Microns.
14 Pre Filter “Pureair” make, EU 4 rating with suitable size.
Efficiency: 90% down to 10 Microns.
15Noise LevelLess than 70 dB ‘A’
16Vibration Level / InstrumentationMinimum
17 Differential Pressure Gauge Magnehelic Gauge – 01 No. (Make: Dwyer, USA) with each unit
(To measure differential pressure across the supply HEPA filter)
18 Supply & Exhaust Motor Blower Assembly Statically & Dynamically balanced with suitable rating and size.
Induction motor with centrifugal blower.
Mounted on unique spring suspension system to reduce vibration and noise.
193 Pins Switch & Socket15 AMPS – 1 set in each unit
20Other SwitchesOn/Off switches for motor blower & CFL lights
21 Microprocessor Control Airflow Display
UV Interlocking
Low Airflow Alarm
Supply & Exhaust Filter Fail Indication

4. Biosafety Cabinet – Class II Type B2

“Pureair” Biosafety Cabinet Class II Type B2 all inflow and downflow air is exhausted after HEPA filtration to the external environment without recirculation within the cabinet. Type B2 cabinets are suitable for work with toxic chemicals used in microbiological processes since no re-circulation occurs. They are among the safest Class II cabinets but difficult to install, balance, and maintain.

1ProductBiosafety Cabinet – Class II Type B2
2Air Balancing0% recirculation, 100% exhaust
3ProtectionProduct, Personnel & Environmental Protection
4Biosafety LevelBSL 1, 2 & 3
5Exhaust Duct RequirementHard Duct Only
6Control PlenumNegative Plenum
7Cleanliness LevelISO Class 5 (ISO 14644-1)
8Average Airflow Velocity Inflow: Minimum 0.45 m/s (90 FPM) at 8” front opening
Downflow: 0.30 m/s (65 FPM)
9Noise Level60 dB(A)
10Light IntensityMinimum 900 lux
11MOC Main Body: GI with PU coated / SS Satin Finish
Working Area: SS 304 Satin Finish
12Supply Filter HEPA EU13, 99.99% @ 0.3 microns
13Exhaust Filter HEPA EU13, 99.99% @ 0.3 microns
14Pre Filter EU4, 90% @ 10 microns
15Noise Level< 70 dB(A)
16Vibration LevelMinimum
17Differential Pressure Gauge Magnehelic Gauge (Dwyer, USA)
Measures HEPA differential pressure
18Motor Blower Assembly Centrifugal blower with spring suspension
19Switch & Socket15A, 3-pin
20SwitchesOn/Off for blower & lights
21Microprocessor Control Airflow display, UV interlock, alarms

5. Biosafety Cabinet – Class III

“Pureair” Biosafety Cabinet Class III ensures absolute safety with fully gas-tight construction. Work is performed using glove ports. The cabinet maintains negative pressure to prevent leaks. HEPA supply and double HEPA exhaust filtration ensure total containment. Suitable for BSL 1–4 including high-risk pathogens.

1ProductBiosafety Cabinet – Class III
2Air Balancing0% recirculation, 100% exhaust
3ProtectionProduct, Personnel & Environmental Protection
4Biosafety LevelBSL 1, 2, 3 & 4
5Exhaust Duct Requirement Inside room / Hard duct only
6Control PlenumNegative Plenum
7Cleanliness LevelISO Class 5 (ISO 14644-1)
8aNegative Pressure > 0.5" WC (closed chamber)
8bFront Sash Airtight with Glove Port
9Noise Level60 dB(A)
10Light IntensityMinimum 900 lux
11MOC Main Body: GI PU coated / SS Satin Finish
Working Area: SS 304 Satin Finish
12Supply Filter HEPA EU13, 99.99% @ 0.3 microns
13Exhaust Filter (Double) Double HEPA EU13, 99.99% @ 0.3 microns
14Pre Filter EU4, 90% @ 10 microns
15Noise Level< 70 dB(A)
16VibrationMinimum
17Differential Pressure Gauge Magnehelic Gauge (Dwyer, USA)
18Motor Blower Assembly Centrifugal blower with spring suspension
19Switch & Socket15A, 3-pin
20Switches On/Off for blower & lights
21Microprocessor Control Airflow display, UV interlock, alarms